Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for...
Xoran Technologies, Inc. began a recall of Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging on . The reason FDA records is "The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0561-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Xoran Technologies, LLC
- Product
- Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.
- Quantity in commerce
- 82
- Distribution
- Nationwide Distribution
- Product codes and lots
- All units.