Z-0561-2010

Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for...

Xoran Technologies, LLC / initiated 2008-09-17 / Terminated / root cause: design

Xoran Technologies, Inc. began a recall of Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging on . The reason FDA records is "The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0561-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Xoran Technologies, LLC
Product
Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
Root cause tag
design
FDA root cause
Device Design
Action
A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.
Quantity in commerce
82
Distribution
Nationwide Distribution
Product codes and lots
All units.