Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray...
Lorad, A Hologic, Inc. began a recall of Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic... on . The reason FDA records is "Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1546-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hologic, Inc.
- Product
- Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
- Product code
- MUE Full Field Digital, System, X-Ray, Mammographic
- Reason for recall
- Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees were notified by an Urgent Recall Notice sent on 3/21/08. The letter informed users that representatives were dispatched to all locations and they have corrected the magnification factor problem. The letter instructed users to re-evaluate any cases read using CAD from the period of installation until the date of system correction. The letter also instructed users that only images read with CAD markers are affected. The letter requests users to return a confirmation of notification. For additional information, contact 1-877-371-4372.
- Quantity in commerce
- 73 units
- Distribution
- Worldwide Distribution including USA and countries of Canada and the Republic of Korea.
- Product codes and lots
- Serial Numbers: 28401060607, 28401060617, 28401060619, 28401060641, 28401060645, 28401071345, 28401071361, 28401071363, 28401082760, 28402050317, 28402071431, 28403060709, 28403071548, 28403071601, 28404060755, 28405060792, 28405060813, 28405071807, 28405071813, 28406071833, 28407050431, 28407060899, 28407071951, 28407072014, 28408050465, 28408072100, 28408072106, 28408072130, 28409030075, 28409030076, 28409050497, 28410072360, 28411061182, 28411061271, 28411072479, 28412050577, 28412050589, 28412061309, 28412061319, 28412061333, 28412061336, 28412061339, 28412072602, 29408072073, 28407072007S, 28409030072RF and 28409030077RM.