Z-1546-2008

Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray...

Hologic, Inc. / initiated 2008-02-28 / Terminated / root cause: software

Lorad, A Hologic, Inc. began a recall of Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic... on . The reason FDA records is "Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1546-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.
Root cause tag
software
FDA root cause
Software design
Action
Consignees were notified by an Urgent Recall Notice sent on 3/21/08. The letter informed users that representatives were dispatched to all locations and they have corrected the magnification factor problem. The letter instructed users to re-evaluate any cases read using CAD from the period of installation until the date of system correction. The letter also instructed users that only images read with CAD markers are affected. The letter requests users to return a confirmation of notification. For additional information, contact 1-877-371-4372.
Quantity in commerce
73 units
Distribution
Worldwide Distribution including USA and countries of Canada and the Republic of Korea.
Product codes and lots
Serial Numbers: 28401060607, 28401060617, 28401060619, 28401060641, 28401060645, 28401071345, 28401071361, 28401071363, 28401082760, 28402050317, 28402071431, 28403060709, 28403071548, 28403071601, 28404060755, 28405060792, 28405060813, 28405071807, 28405071813, 28406071833, 28407050431, 28407060899, 28407071951, 28407072014, 28408050465, 28408072100, 28408072106, 28408072130, 28409030075, 28409030076, 28409050497, 28410072360, 28411061182, 28411061271, 28411072479, 28412050577, 28412050589, 28412061309, 28412061319, 28412061333, 28412061336, 28412061339, 28412072602, 29408072073, 28407072007S, 28409030072RF and 28409030077RM.