Z-1501-2008

Zimmer Hemovac Wound Drainage Device, 400 ml COMP EVAC 3/32" PVC w/trocar, 2,4mm, 7Fr, PVC, 107cm, small...

Zimmer GmbH / initiated 2008-01-29 / Terminated / root cause: packaging

Zimmer Inc. began a recall of Zimmer Hemovac Wound Drainage Device, 400 ml COMP EVAC 3/32" PVC w/trocar, 2,4mm, 7Fr, PVC, 107cm, small, sterile, latex free, Rx only; Zimmer, Swindon... on . The reason FDA records is "The sterility of the device may be compromised due to the possible presence of slits or pinholes in the packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1501-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Hemovac Wound Drainage Device, 400 ml COMP EVAC 3/32" PVC w/trocar, 2,4mm, 7Fr, PVC, 107cm, small, sterile, latex free, Rx only; Zimmer, Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio; REF 00-2595-000-10.
Product code
GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
Reason for recall
The sterility of the device may be compromised due to the possible presence of slits or pinholes in the packaging.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Consignees were notified via recall letter dated 1/29/08 to locate and return the affected product. The recall action was expanded to additional products/lots via recall letter dated 3/17/08. Custom kit packers were instructed to notify their customers.
Distribution
Nationwide, Australia, Canada, Chile, China, Czech Republic, Dubai, Ecuador, France, Germany, Greece, Hungary, Italy, Japan, Korea, Poland, Romania, Russia, Saudi Arabia, Serbia & Montenegro, Singapore, Slovakia, South Africa, Spain, and Switzerland.
Product codes and lots
All lots manufactured on or before 1/18/08. Product has a 5 year expiration period.