Z-1588-2008

VARIANT -thalassemia Short Program Hemoglobin Testing System, Model Number 270-0019 (250 tests), with...

Bio-Rad Laboratories Inc. / initiated 2008-03-18 / Terminated

Bio-Rad Laboratories Inc began a recall of VARIANT -thalassemia Short Program Hemoglobin Testing System, Model Number 270-0019 (250 tests), with -thalassemia ROM card lot PA 62600, PIN 2700104 on . The reason FDA records is "Some product ROM cards have an incorrect program loaded and are unable to be updated". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1588-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
VARIANT -thalassemia Short Program Hemoglobin Testing System, Model Number 270-0019 (250 tests), with -thalassemia ROM card lot PA 62600, PIN 2700104
Product code
JPD Hemoglobin A2 Quantitation
Reason for recall
Some product ROM cards have an incorrect program loaded and are unable to be updated.
FDA root cause
Mixed-up of materials/components
Action
The firm initiated its recall on March 18, 2008 by issuing a recall notice to customers and subsidiaries via e-mail. Not all VARIANT -Thalassemia Short ROM Cards are affected. Inventory should be examined and quarantine the product subject to this Recall. Contact your regional Bio-Rad office for ROM Card replacement or if you have questions.
Quantity in commerce
64 kits total
Distribution
Product was distributed to 17 consignees and 7 subsidiaries Worldwide. Distribution included to states of CA, TX, WI, VA, UT, PA, TN, OH, FL, MA, OR, AR, IL, MD, CO, and NY, and countries of UK, Australia, Canada, France, Singapore, New Zealand, Italy.
Product codes and lots
Lot number 70080440