InCompass Spinal Fixation System T-Handle Driver; Part #2151-4 (9.5" shaft) and 2151-5 (13" shaft)...
Abbott Spine began a recall of InCompass Spinal Fixation System T-Handle Driver; Part #2151-4 (9.5" shaft) and 2151-5 (13" shaft); Non-sterile, Reusable Instrument; manufactured by Abbott... on . The reason FDA records is "Reports received regarding disassembly of T-Handle Driver components during surgery dropping the closure top into the surgical site". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1592-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Spinal Concepts, Inc.
- Product
- InCompass Spinal Fixation System T-Handle Driver; Part #2151-4 (9.5" shaft) and 2151-5 (13" shaft); Non-sterile, Reusable Instrument; manufactured by Abbott Spine, Austin, TX 78727.
- Product code
- MNI Orthosis, Spinal Pedicle Fixation
- Reason for recall
- Reports received regarding disassembly of T-Handle Driver components during surgery dropping the closure top into the surgical site.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Firm sent notification letters to sales agents via letter on 03/24/08. Enclosed with the notification letter, the firm provided updated product inserts for the InCompass and PathFinder Spinal Fixation System (Cat. #2199-0002-LB Rev. H and #1199-0002-LB Rev. H, respectively), as well as the "Disassembly and Assembly Instructions for PathFinder & InCompass T-Handle Drivers" (Cat #1199-0040-MKC Rev. A). Sales Agents instructed to distribute updated inserts and instructions to all locations that handle the drivers for cleaning and to destroy all copies of old inserts. If you have questions or have not received the recall notice from the firm, contact Kirstin Widding at 512-533-1852.
- Quantity in commerce
- 278 units.
- Distribution
- Nationwide Distribution --- including states of AR, AZ, CA, CT, CO, FL, IL, LA, MD, MI, MO, MN, MT, NC, NM, NJ, NY, OH, OK, OR, PA, SD, TX, VA and WI.
- Product codes and lots
- All lots.