Logix TPN Software Solution, catalog 2M8410, consists of the Logix Order Entry (OE) Installation CD, version...
Baxter Healthcare Corp. began a recall of Logix TPN Software Solution, catalog 2M8410, consists of the Logix Order Entry (OE) Installation CD, version 1.0.28 and Logix Compounder Manager (CM)... on . The reason FDA records is "Software anomaly results in inaccurate information being printed out on the Delivery Report, even though the compounder performed the compounding correctly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1640-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Logix TPN Software Solution, catalog 2M8410, consists of the Logix Order Entry (OE) Installation CD, version 1.0.28 and Logix Compounder Manager (CM) Installation CD, version 3.1.8. The software is included in a number of Baxter's pharmacy compounders. The software assists the pharmacist in management of compounding, creating and authorizing nutritional solutions intended for IV administration.
- Product code
- LHI Set, I.V. Fluid Transfer
- Reason for recall
- Software anomaly results in inaccurate information being printed out on the Delivery Report, even though the compounder performed the compounding correctly.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Baxter sent the 3/28/08 Urgent Device Correction letter to affected customers, to the attention of the Director of Pharmacy, via first class mail to advise them of the potential inaccuracy of the Delivery Report due to a software anomaly. The accounts were instructed to review each delivery report for accuracy. If the report contains zero actual volumes the user must manually reprint the report. The letter outlined the process for reprinting the Delivery Report. Any questions were directed to the Baxter Nutritional Professional Services at 1-800-422-2751. The accounts were requested complete the enclosed reply form acknowledging receipt and understanding of the letter and dissemination of the information to their staff. A software upgrade will be provided. If you have questions about this letter or you did not receive this letter contact Baxter Nutrition Professional Services at 1-800-422-2751.
- Quantity in commerce
- 5 units
- Distribution
- Arizona, Minnesota, Wisconsin, Puerto Rico and Canada
- Product codes and lots
- Software version 1.0.2 (comprised of Logix OE version 1.0.28 and Logix CM version 3.1.8)