Z-0995-2009
Cyberonics VNS Therapy Demipulse Duo Generator, Model 104; an implantable, multiprogrammable, vagus nerve...
LivaNova USA, Inc. / initiated 2008-02-07 / Terminated / root cause: software
Cyberonics, Inc began a recall of Cyberonics VNS Therapy Demipulse Duo Generator, Model 104; an implantable, multiprogrammable, vagus nerve stimulator on . The reason FDA records is "Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0995-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Cyberonics VNS Therapy Demipulse Duo Generator, Model 104; an implantable, multiprogrammable, vagus nerve stimulator.
- Product code
- LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
- Reason for recall
- Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Cyberonics notified consignees of recall via letter on 02/07/08. Consignees were notified of reset issue in which during a strong magnetic field, the software may report an error even though no error occurred. This error causes the generator to reset, which disables the generator (i.e., no stimulation delivered). Physicians should inform any potential and currently implanted patients of the possibility of this event. Patients should continue with their regularly scheduled visits as determined by their physician. If this event is suspected, physicians should interrogate the device as described in the Physicians Manual to confirm the generator has been reset and Cyberonics should be contacted for any additional instruction.
- Quantity in commerce
- 108 units.
- Distribution
- Worldwide Distribution - USA, Belgium, Germany, Norway, Spain, Sweden, the Netherlands, and United Kingdom.
- Product codes and lots
- All lots.