Z-0868-2008

Roche/Hitachi Modular Analytics System, Modular E Module GMMI immunoassay analyzer; Catalog numbers...

Roche Diagnostics GmbH / initiated 2008-03-21 / Terminated

Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular Analytics System, Modular E Module GMMI immunoassay analyzer; Catalog numbers 04998642001 and 03617505001. Intended for qualitative and... on . The reason FDA records is "Possible mismatch between patient and result". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0868-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche/Hitachi Modular Analytics System, Modular E Module GMMI immunoassay analyzer; Catalog numbers 04998642001 and 03617505001. Intended for qualitative and quantitative in vitro determinations using a wide variety of tests. Roche Diagnostics Corporation, 9115 Hague Road, Indianapolis, IN, 46250
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Possible mismatch between patient and result.
FDA root cause
Process change control
Action
Consignees were notified via Urgent Medical Device Correction letter dated 3/21/08, which instructed them to not manually order QC test selections while the system is in operation mode and that they can only manually order QC when the system is in Stand-by mode. Consignees are instructed to consult with their physicians or pathologist to determine specific clinical implications for their patients. The firm states that this issue will be resolved in the next version of software.
Quantity in commerce
11 of all models.
Distribution
Nationwide Distribution --- including states of Florida, Georgia, Indiana, New Jersey and North Carolina.
Product codes and lots
All units with software version 08-02.