Z-0869-2008
Roche/Hitachi Modular Analytics System, Modular D Module GMMI clinical chemistry analyzer; Catalog number...
Roche Diagnostics GmbH / initiated 2008-03-21 / Terminated
Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular Analytics System, Modular D Module GMMI clinical chemistry analyzer; Catalog number 03284522001. Intended for qualitative and... on . The reason FDA records is "Possible mismatch between patient and result". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0869-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi Modular Analytics System, Modular D Module GMMI clinical chemistry analyzer; Catalog number 03284522001. Intended for qualitative and quantitative in vitro determinations using a wide variety of tests. Roche Diagnostics Corporation, 9115 Hague Road, Indianapolis, IN, 46250
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Possible mismatch between patient and result.
- FDA root cause
- Process change control
- Action
- Consignees were notified via Urgent Medical Device Correction letter dated 3/21/08, which instructed them to not manually order QC test selections while the system is in operation mode and that they can only manually order QC when the system is in Stand-by mode. Consignees are instructed to consult with their physicians or pathologist to determine specific clinical implications for their patients. The firm states that this issue will be resolved in the next version of software.
- Quantity in commerce
- 11 of all models
- Distribution
- Nationwide Distribution --- including states of Florida, Georgia, Indiana, New Jersey and North Carolina.
- Product codes and lots
- All units with software version 08-02.