HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence...
Focus Diagnostics Inc began a recall of HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2... on . The reason FDA records is "A small percentage of samples tested in specific lots, the HSV common antigen band appears lighter (less reactive) than the reading control band. Reduced reactivity of the common antigen band may lead to inaccurate...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1764-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DiaSorin Molecular LLC
- Product
- HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera.
- Product code
- LGC Enzyme Linked Immunoabsorbent Assay, Herpes Simplex Virus, Non-Specific
- Reason for recall
- A small percentage of samples tested in specific lots, the HSV common antigen band appears lighter (less reactive) than the reading control band. Reduced reactivity of the common antigen band may lead to inaccurate interpretation of a sample as negative.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Focus notified all clients which received the recalled lots starting on July 6, 2007 via an Important Medical Device Information letter by fax or email. The letter explains how to identify suspect results and the steps to be taken if suspect results are obtained. If customers find the interpretation of results with the indicated kit lots to be problematic, customers have the option of having their kits replaced at no charge within 90 days of receiving the notification.
- Quantity in commerce
- Lot Code 063169 - 97 Kits, Lot Code 063354 - 16 Kits, Lot Code 063355 - 74 Kits
- Distribution
- Worldwide Distribution --- USA including states of OH, CA, CT, IL, NY, and countries of Barbados, Canada, Germany, Israel, Slovakia, France, and England.
- Product codes and lots
- Lot Codes 063169 (Expires December 2007), 063354 (Expires October 2007), 063355 (Expires July 2007)