24mm Rotating C-Mount Coupler, Model Number 1088-020-121S2. The device in indicated for use in general...
Stryker Endoscopy began a recall of 24mm Rotating C-Mount Coupler, Model Number 1088-020-121S2. The device in indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy... on . The reason FDA records is "The coupler on this device was manufactured without set-screws that are used to hold and center the rear assembly and the endobody. The coupler is part of the camera system and affects the ability to focus". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1558-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- 24mm Rotating C-Mount Coupler, Model Number 1088-020-121S2. The device in indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope or arthroscope is indicated for use.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- The coupler on this device was manufactured without set-screws that are used to hold and center the rear assembly and the endobody. The coupler is part of the camera system and affects the ability to focus.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Only one customer had all of the defective couplers. The firm called this customer on March 27, 2008, and asked that the 28 couplers remaining in their possession be returned.
- Quantity in commerce
- 28 units
- Distribution
- Product was delivered to one consignee in Iowa.
- Product codes and lots
- Model Number: 1088-020-121S2 Lot Numbers: Shipped: 07L046134, 07L046144, 07L046154, 07L046164, 07L046184, 07L046194, 07L046214, 07L046224, 07L046234, 07L046244, 07L046254, 07L046274, 07L046284, 07L046294, 07L046304, 07L046324, 07L046334, 07L046344, 07L046364, 07L046374, 07L046384, 07L046394, 07L046404, 07L046414, 07L046424, 07L046454, 07L046464, and 07L046474.