Z-1603-2008
Hill-Rom 100 Low Bed; Model No. P3930
Hill-Rom Inc. / initiated 2008-04-08 / Terminated / root cause: design
Hill-Rom, Inc. began a recall of Hill-Rom 100 Low Bed; Model No. P3930 on . The reason FDA records is "The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1603-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom 100 Low Bed; Model No. P3930.
- Product code
- FNL Bed, Ac-Powered Adjustable Hospital
- Reason for recall
- The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified via letter dated April 8, 2008 to lower these beds to their lowest position, to remove the beds from use if they move more than 1/2 inch with the brakes locked and that the firm will repair these beds when fixes become available. If you have questions regarding this recall, contact Hill-Rom Technical Support at 800-445-3730.
- Quantity in commerce
- 753 beds
- Distribution
- Worldwide: USA and Canada.
- Product codes and lots
- All units.