Z-1374-2008

Philips OmniDiagnost Eleva X-Ray System is a diagnostic imaging device intended for radiographic...

Philips North America LLC / initiated 2008-03-22 / Terminated / root cause: software

Philips Medical Systems North America Co. Phillips began a recall of Philips OmniDiagnost Eleva X-Ray System is a diagnostic imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications on . The reason FDA records is "Unexpected movement of the X-ray table may occur. These movements may lead to patient injury, e.g., an uncontrolled tilt may cause the patient to slip off". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1374-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips OmniDiagnost Eleva X-Ray System is a diagnostic imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Unexpected movement of the X-ray table may occur. These movements may lead to patient injury, e.g., an uncontrolled tilt may cause the patient to slip off.
Root cause tag
software
FDA root cause
Software design
Action
On 3/22/08, the firm mailed letters to their customers via certified mail. The letter informed the customers of the recall, advising the customers to, "Use the emergency stop button or release the initiated movement command to halt undesired table movement. If you perform digital and conventional acquisition, you can avoid the problem by closing the single examinations daily or by clearing the patient database regularly." The letter further informed the customers that a Philips representative will be installing updated software at a future date. Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.
Quantity in commerce
22 units
Distribution
Nationwide Distribution -- including states of CA, CO, KY, TN, LA, MI, FL, MS, TX, VA, WI, MO, AZ, and NY.
Product codes and lots
Product number 9896 002 04432. SN number: 43, 49, 52, 107, 123, 131, 152, 157, 171, 179, 185, 202, 266, 333, 337, 389, 408, 446, 485, 491, 514, 515; Site ID #: 506241, 500046, 521648, 522220, 520020, 533112, 533314, 533471, 533530, 535257, 535118, 535924, 540860, 541490, 539106, 547017, 547142, 549798, 547348, 550258, 553676, and 553235.