Z-1692-2008

Puritan-Bennett 800 Series Ventilator Backup Power Source (BPS) Model 802, Puritan-Bennett Corporation, A...

Medtronic, Inc. / initiated 2008-03-21 / Terminated / root cause: manufacturing process

Covidien Limited began a recall of Puritan-Bennett 800 Series Ventilator Backup Power Source (BPS) Model 802, Puritan-Bennett Corporation, A subsidiary of Mallinckrodt Inc., Carlsbad, CA on . The reason FDA records is "Wiring in battery back-up power supply may short and cause thermal damage to ventilator". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1692-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Puritan-Bennett 800 Series Ventilator Backup Power Source (BPS) Model 802, Puritan-Bennett Corporation, A subsidiary of Mallinckrodt Inc., Carlsbad, CA
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Wiring in battery back-up power supply may short and cause thermal damage to ventilator.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
All customers were notified by Federal Express letter and were contacted by phone through field service engineers. They were told to examine their inventory for any installed or spare part battery assemblies and to remove and quarantine these parts. Any ventilators with installed affected battery assemblies were to be taken out of service.
Quantity in commerce
6 Ventilators and 151 factory replacement units
Distribution
Nationwide, including one VA facility in IL and two military facilities in NC and NY. No foreign distribution.
Product codes and lots
4-070523-SP Battery Assembly 4-840120DIUU-US 4-070520-SP BPS