Z-1704-2008

Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is...

Biosense Webster Inc / initiated 2008-03-25 / Terminated / root cause: design

Biosense Webster Inc began a recall of Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is indicated for multiple electrode electrophysiological... on . The reason FDA records is "This recall was initiated after the firm received several complaints reporting issues with deflection. The catheter may become caught in the mitral valve, requiring surgical procedures, valve may be torn while trying to...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1704-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biosense Webster Inc
Product
Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e., recording and stimulation only. The device is designed to obtain electrograms in the atrial regions of the heart. 20 Poles, US Catalog Number D7L202515RT, EU Catalog Number 35T2515R
Product code
DRF Catheter, Electrode Recording, Or Probe, Electrode Recording
Reason for recall
This recall was initiated after the firm received several complaints reporting issues with deflection. The catheter may become caught in the mitral valve, requiring surgical procedures, valve may be torn while trying to dislodge, unable to retract the catheter through the sheath and require surgical intervention to remove, pulling on catheter to remove may lead to large atrial septal tear, and sep
Root cause tag
design
FDA root cause
Device Design
Action
A letter entitled, "Urgent Safety Field Notice" dated March 25, 2008, and certification form were sent to all customers and Johnson & Johnson affiliates who purchased the product in the past year as of March 25, 2008, due to the expiration date. For customers in the US, two letters and forms were sent out - one addressed to the EP/Cardiology Lab and a second to the Risk Management Office. If you have questions or have not received a letter contact Customer Service at 1-800-729-9010.
Quantity in commerce
14,295 units for both products
Distribution
Nationwide and Worldwide to: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czechia, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary , Iran, Ireland, Isra¿¿l, Italy, Japan, Lithuania, Luxembourg, Mexico, Middle East, Netherlands, Norway, Panama, Poland, Portugal, Russia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom
Product codes and lots
All lots