Z-1733-2008
Zimmer Spine Cyclone anterior cervical plate, 3 Level, 45MM (Ti-6Al-4V); REF 600-03-145
Zimmer GmbH / initiated 2008-03-26 / Terminated / root cause: design
Zimmer Inc. began a recall of Zimmer Spine Cyclone anterior cervical plate, 3 Level, 45MM (Ti-6Al-4V); REF 600-03-145 on . The reason FDA records is "The screw locking cap may fracture when the surgeon rotates it into the locked position". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1733-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Spine Cyclone anterior cervical plate, 3 Level, 45MM (Ti-6Al-4V); REF 600-03-145.
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- The screw locking cap may fracture when the surgeon rotates it into the locked position.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified via voicemail on 3/26/08 and a letter was sent to all distributors on 3/27/08 informing them that the plate should no longer be used and that all lots are being recalled. A letter dated 6/18/08 was issued to implant physicians recommending patient monitoring. Contact Zimmer Spine at 1-952-857-5636 for assistance.
- Distribution
- Arizona, California, Florida, Georgia, Louisiana, Michigan, Minnesota, Montana, New Hampshire, North Carolina, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, Washington and Wisconsin.
- Product codes and lots
- All lots.