Z-1738-2008

Zimmer Spine Cyclone anterior cervical plate, 3 Level, 60MM (Ti-6Al-4V); REF 600-03-160

Zimmer GmbH / initiated 2008-03-26 / Terminated / root cause: design

Zimmer Inc. began a recall of Zimmer Spine Cyclone anterior cervical plate, 3 Level, 60MM (Ti-6Al-4V); REF 600-03-160 on . The reason FDA records is "The screw locking cap may fracture when the surgeon rotates it into the locked position". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1738-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Spine Cyclone anterior cervical plate, 3 Level, 60MM (Ti-6Al-4V); REF 600-03-160.
Product code
KWQ Appliance, Fixation, Spinal Intervertebral Body
Reason for recall
The screw locking cap may fracture when the surgeon rotates it into the locked position.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified via voicemail on 3/26/08 and a letter was sent to all distributors on 3/27/08 informing them that the plate should no longer be used and that all lots are being recalled. A letter dated 6/18/08 was issued to implant physicians recommending patient monitoring. Contact Zimmer Spine at 1-952-857-5636 for assistance.
Distribution
Arizona, California, Florida, Georgia, Louisiana, Michigan, Minnesota, Montana, New Hampshire, North Carolina, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, Washington and Wisconsin.
Product codes and lots
All lots.