Essenta DR - digital multifunctional x-ray system (application SW release 1.0.2 and lower). The device is...
Philips Medical Systems North America Co. Phillips began a recall of Essenta DR - digital multifunctional x-ray system (application SW release 1.0.2 and lower). The device is intended for routine exams, including specialist... on . The reason FDA records is "Bright artifacts may appear on a patient image from a previous exposure, which might lead to a misdiagnosis". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1895-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Essenta DR - digital multifunctional x-ray system (application SW release 1.0.2 and lower). The device is intended for routine exams, including specialist areas like trauma or pediatric work, excluding mammography.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Bright artifacts may appear on a patient image from a previous exposure, which might lead to a misdiagnosis.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 4/08/08, the firm notified customers by letter, sent by certified mail. The letter explained the recall and advised customers of the following: "If direct radiation reaches the detector, the collimation should be not too tight, such that the area of direct radiation is larger than 30 mm x 30 mm. (1.2 inch by 1.2 inch). The reason is that a minimum area is needed to detect and prevent memory artifact... All possibly affected products in the field will be upgraded with FCO 71200026 to resolve the issue." Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.
- Quantity in commerce
- 10 units
- Distribution
- Nationwide Distribution -- including states of SC, MN, CA, GA, TX, NY, OR and UT.
- Product codes and lots
- Serial Number: 368645, 372631, 375371/SN0800002, 375730/SN07000032, 375372/SN07000028, 383083/SN07000023, 378493/SN07000025, 379442/SN0800001, 379183/SN07000030 and 379050/SN0800005; Equip No: 553491, 554498, 554944, 554358, 553270, 557203, 556089, 556080, 555457 and 556275.