Z-1903-2008

IsoMed Refill Kit, Model 8553, Lot Number 60538731, is intended for use in refilling Medtronic IsoMed pumps...

Medtronic Neuromodulation / initiated 2008-02-04 / Terminated / root cause: manufacturing process

Medtronic Neuromodulation began a recall of IsoMed Refill Kit, Model 8553, Lot Number 60538731, is intended for use in refilling Medtronic IsoMed pumps. Contents: Two 22-gauge needles, extension tubing... on . The reason FDA records is "Medtronic is retrieving all unused packages of ISOMED Refill Kit Model 8553, Lot 60538731. This lot of product was released using incorrect endotoxin specification limits. Endotoxin contamination can potentially lead to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1903-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
IsoMed Refill Kit, Model 8553, Lot Number 60538731, is intended for use in refilling Medtronic IsoMed pumps. Contents: Two 22-gauge needles, extension tubing set with y-connector and clamp, Filter, 10-ml syringe, 60-ml lidded syringe, Fenestrated drape, Template; Use by Date May 10, 2009; Medtronic, Inc., Minneapolis, MN. Product is packaged in a sterile container.
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Medtronic is retrieving all unused packages of ISOMED Refill Kit Model 8553, Lot 60538731. This lot of product was released using incorrect endotoxin specification limits. Endotoxin contamination can potentially lead to fever, malaise, respiratory distress, septic shock, and death.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by Medtronic field personnel, either by personal visits or telephone, beginning 2/4/08. This included the 8 additional consignees located outside of the U.S. Medtronic advised customers not to use any unused packages in their inventory and that a field representative would assist them in returning the product for replacement and/or credit. Upon completion of each follow-up, the field representatives completed and returned a Reply Card documenting the notification. If you have questions or comments, contact Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Quantity in commerce
1982 kits
Distribution
Worldwide -- including the U.S. (AK, AL, AR, AZ, CA, CT, DC,DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and WV) and Argentina, Australia, and Hong Kong.
Product codes and lots
Lot 60538731, Use by Date is May 10, 2009.