Bariatric Wheelchair, 28" Wide and 20" Deep Seat, 700 lb. Weight Capacity; Manufactured in Taiwan for Medline...
Medline Industries Inc began a recall of Bariatric Wheelchair, 28" Wide and 20" Deep Seat, 700 lb. Weight Capacity; Manufactured in Taiwan for Medline Industries, Inc., Mundelein, IL 60060 USA; Item... on . The reason FDA records is "Injuries to fingers. There is a potential for users to pinch their fingers between the seat bars when opening the wheelchairs. The injuries can be severe from fracture to severing". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1772-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Bariatric Wheelchair, 28" Wide and 20" Deep Seat, 700 lb. Weight Capacity; Manufactured in Taiwan for Medline Industries, Inc., Mundelein, IL 60060 USA; Item #MDS809800 comes with Removable Desk-Length Arms, Swing-Away Detachable Footrests, Hemi-Height Adjustable. item #MDS809850 comes with Removable Desk-Length Arms, Swing-Away Detachable Elevating Legrests, Hemi-Height Adjustable.
- Product code
- IOR Wheelchair, Mechanical
- Reason for recall
- Injuries to fingers. There is a potential for users to pinch their fingers between the seat bars when opening the wheelchairs. The injuries can be severe from fracture to severing.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- A Voluntary Field Correction letter dated 3/31/08 was issued to customers. The letter informed them of reports of users catching their fingers under the seat bars, which are identified on the chair with warning stickers. The customers were requested to change the backrest with the enclosed replacement upholstery kit, properly dispose of the old back rest, and place the enclosed cautionary tags on both sides of the chair seat. Instructions for installation of the upholstery and cautionary tags were included, as was a sheet entitled "How to Open a Wheelchair" which gives step by step instructions with illustrations for opening the wheelchair, as well as the warning "CAUTION KEEP YOUR FINGERS ON TOP OF THE UPHOLSTERY AT ALL TIMES. DO NOT PLACE YOUR FINGERS ON THE SIDES OR BOTTOM OF THE CHAIR SEAT." The accounts were requested to complete the enclosed response sheet and fax it back to Medline at 847-643-4482, indicating the serial numbers of chairs corrected. Any questions were directed to 800-633-5463, extension 5349, between 8:00 AM and 5:00 PM CST. Distributors were requested to notify their accounts of the recall.
- Quantity in commerce
- 201 wheelchairs
- Distribution
- Class II Recall - Worldwide Distribution --- including USA and Canada.
- Product codes and lots
- Item numbers MDS809800 and MDS809850, all wheelchairs purchased between 12/01/2005 and 08/31/2007