Z-2139-2008

CELL-DYN Sapphire Hematology Analyzer, List number 08H00-01, LIS/ Middleware Interface Specification, Abbott...

Abbott Medical / initiated 2008-04-11 / Terminated / root cause: software

Abbott Laboratories began a recall of CELL-DYN Sapphire Hematology Analyzer, List number 08H00-01, LIS/ Middleware Interface Specification, Abbott Diagnostics Division, Santa Clara, CA For in vitro... on . The reason FDA records is "Table in LIS Specification depicting the association between Record ID and Numerical Result Label is incorrect. If a record ID is used to configure the system for mapping, results will come out nonsensical". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2139-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
CELL-DYN Sapphire Hematology Analyzer, List number 08H00-01, LIS/ Middleware Interface Specification, Abbott Diagnostics Division, Santa Clara, CA For in vitro diagnostic use in counting and characterizing blood cells.
Product code
GKZ Counter, Differential Cell
Reason for recall
Table in LIS Specification depicting the association between Record ID and Numerical Result Label is incorrect. If a record ID is used to configure the system for mapping, results will come out nonsensical.
Root cause tag
software
FDA root cause
Software change control
Action
Recall initiated on April 11, 2008. A Product Correction letter and customer reply form were sent to all currently active CELL-DYN Sapphire Customers. The letter requests that users be certain that the transmission of test results receipt by the host interface is mapped according to the Numerical Result Label, when configuring their CELL-DYN Sapphire Hematology Analyzer for results transmission to a Clinical Laboratory Information System or Middleware. Affected product was distributed to a total of 4 distributors. However, notification was made to all 154 customers that received the CELL-DYNE Sapphire product. If the customer or the health care providers that they serve have any questions regarding this information, U.S. customers should call Customer Support at 1-877-4ABBOTT.
Quantity in commerce
15 copies of the specification were sent worldwide
Distribution
Nationwide Distribution.
Product codes and lots
Revision B.