Z-1778-2008

Xcelerate Patella System Reamer Shaft Assembly; Non-Sterile; Howmedica Osteonics Corp., 325 Corporate Drive...

Howmedica Osteonics Corp / initiated 2005-04-15 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Xcelerate Patella System Reamer Shaft Assembly; Non-Sterile; Howmedica Osteonics Corp., 325 Corporate Drive; Mahwah, NJ 07430 Authorized representative in... on . The reason FDA records is "The Xcelerate Patella Reamer Shaft Assembly and the Reamer Adapter do not mate properly. This could result in the removal of excessive bone from the patella leading to a thinner patella; a patella that does not have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1778-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Xcelerate Patella System Reamer Shaft Assembly; Non-Sterile; Howmedica Osteonics Corp., 325 Corporate Drive; Mahwah, NJ 07430 Authorized representative in Europe: Stryker France ZAC Satolas Green Pusignan Cedex, France
Product code
HTO Reamer
Reason for recall
The Xcelerate Patella Reamer Shaft Assembly and the Reamer Adapter do not mate properly. This could result in the removal of excessive bone from the patella leading to a thinner patella; a patella that does not have enough bone for adequate engagement with pegs or patella fracture.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An IMPORTANT MARKET WITHDRAWAL letter was issued on April 19, 2005 and sent Federal Express to those consignees who received the device. The letter instructed them to examine their inventory to identify instruments, reconcile these instruments on an attached accountability form; fax a copy of this form within 2 days of receipt of the letter; retrieve and return affected product; contact Stryker Customer Service to re-order.
Quantity in commerce
24 total
Distribution
Class II Recall - Worldwide Distribution --- including USA and Canada, United Kingdom.
Product codes and lots
Product no. 8001-1702; Lot codes: MCYL30A, and MCYL30B.