PathFinder Cannulated T-Handle Bone Awl, Part Number 1155-2; distributed by Abbott Spine, Inc., Austin, TX...

Spinal Concepts, Inc. / initiated 2008-04-17 / Terminated / root cause: design

Abbott Spine began a recall of PathFinder Cannulated T-Handle Bone Awl, Part Number 1155-2; distributed by Abbott Spine, Inc., Austin, TX 78727 on . The reason FDA records is "Breakage - Tips on the Pathfinder Bone Awl may bend or break prematurely. The firm has received internal non-conformances and complaints where the tips of the Awl were worn and damaged. Of the tip breaks, six have had...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1784-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Spinal Concepts, Inc.
Product
PathFinder Cannulated T-Handle Bone Awl, Part Number 1155-2; distributed by Abbott Spine, Inc., Austin, TX 78727.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Breakage - Tips on the Pathfinder Bone Awl may bend or break prematurely. The firm has received internal non-conformances and complaints where the tips of the Awl were worn and damaged. Of the tip breaks, six have had material missing. During the same period, there were 15 complaints reported where the tip broke off at the weld after repeated use (1-3 years). The firm found several possible ca
Root cause tag
design
FDA root cause
Device Design
Action
Phone calls and emails were used to disseminate information about this recall. Firm notified consignees via telephone beginning on 04/17/08. Consignees were told that some problems were found through normal wear and tear as well as product mis-use and material weaknesses. The firm explained that they noticed from wear and tear from use as well as damage from use could cause the tip to become dull and flared causing breakage. The firm explained that they redesigned the awl and requested that consignees return outstanding inventory, which will be replaced with a new version of the awl.
Quantity in commerce
180 units.
Distribution
Class II Recall - Worldwide Distribution --- USA including states of AR, AZ, CA, CT, CO, FL, IL, MD, MI, MO, MT, NC, NJ, NV, NY, NM, OH, OK, PA, SD, TX and WI, and countries of Argentina, Australia, Belgium, Canada, China, Columbia, Cyprus, Germany, Greece, Italy, Japan, Mexico, Thailand, Turkey and UK/Ireland.
Product codes and lots
All lots.