Z-1798-2008
Contour Genesis Inf/IR Tubing; Part Number CG-UT; distributed by Mentor Texas, LP, Irving, TX 75038
Mentor / initiated 2008-04-21 / Terminated / root cause: packaging
Mentor Texas, Inc began a recall of Contour Genesis Inf/IR Tubing; Part Number CG-UT; distributed by Mentor Texas, LP, Irving, TX 75038 on . The reason FDA records is "During packaging integrity validation, it was found that various product packages could fail, compromising the sterility of the devices". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1798-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mentor
- Product
- Contour Genesis Inf/IR Tubing; Part Number CG-UT; distributed by Mentor Texas, LP, Irving, TX 75038.
- Product code
- FPK Tubing, Fluid Delivery
- Reason for recall
- During packaging integrity validation, it was found that various product packages could fail, compromising the sterility of the devices.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- A Customer Letter was issued to all consignees by the firm on 4/21/2008. Consignees were informed that a recent testing of these products discovered a potential defect in the packaging configuration, which could result in a loss of package integrity. Consignees were asked to review inventory and dispose of product and contact firm for credit.
- Quantity in commerce
- 3,359 boxes.
- Distribution
- Class II Recall - Worldwide Distribution ---- including USA, Puerto Rico and countries of Argentina, Austria, Bermuda, Brazil, Canada, China, Costa Rica, Cyprus, Denmark, England, Germany, Greece, Guatemala, India, Italy, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Morocco, Pakistan, Panama, Paraguay, Portugal, Slovak Republic, Slovenia, Spain, South Africa, South Korea, Switzerland, Taiwan, Thailand, Turkey and Venezuela.
- Product codes and lots
- Lot #: CY5906, CY5912, CY5920, CY5924, CY5932, CY5938, CY5944, CY5952, CY5958, CY5974, CY5976, CY6986, CY6990, CY6996, CY6102, CY6110, CY6114, CY6126, CY6148, CY6154, CY6170, CY7184, CY7188, CY7190, CY7206, CY7218, CY7226, CY7240, CY7252, CY7264, CY7278, CY8288 and CY8316.