Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074. All...

Philips North America LLC / initiated 2008-04-30 / Terminated / root cause: labeling and UDI

Philips Medical Systems began a recall of Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074. All four model numbers have the same user manual. The... on . The reason FDA records is "Some users of SureSigns VS3 Signs Monitor may have assumed that the SpO2 Non-Pulsatile and SpO2 No Sensor technical alarms were enabled when the SureSigns VS3 Vital Signs Monitor was not in Interval non-invasive blood...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1911-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074. All four model numbers have the same user manual. The device is intended for monitoring, recording and alarming of multiple physiological parameters.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Some users of SureSigns VS3 Signs Monitor may have assumed that the SpO2 Non-Pulsatile and SpO2 No Sensor technical alarms were enabled when the SureSigns VS3 Vital Signs Monitor was not in Interval non-invasive blood pressure (NBP) mode. These alarms are only enabled when interval NBP mode is selected. If when monitoring SpO2, the clinician does not notice when the sensor is no longer attached, the patient is not being monitored through Spo2.This could result in delay or lack of needed therapy. The labeling provided with the monitor did not fully describe this feature.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
All affected customers will receive the FSN (Field Safety Notice )informing them about the labeling correction. Included with the FSN will be an addendum to the instructions for use. customers are asked to read the addendum and insert a copy into each of their copies of the VS3 instructions for use. Customers will also receive a Medical Device Correction Confirmation Card that they will be asked to fill out and fax back to their local Philips representative. Contact your local Philips support organization at 1-800-722-9377 in the United States or 1-800-323-2280 in Canada.
Quantity in commerce
7,529 units
Distribution
Worldwide Distribution to United States, Canada and Foreign countries
Product codes and lots
SureSigns VS3; Model numbers 863071, 863072, 863073, 863074