Z-1698-2008

Urethral Warming System-Galil Medical Warming Catheter. The product is contained within prostate kits. This...

Galil Medical Ltd. / initiated 2007-12-11 / Terminated / root cause: manufacturing process

Galil Medical began a recall of Urethral Warming System-Galil Medical Warming Catheter. The product is contained within prostate kits. This product is used in cryosurgery to prevent freezing... on . The reason FDA records is "The warming catheter was manufactured with the inlet and outlet tubes switched". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1698-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Galil Medical Ltd.
Product
Urethral Warming System-Galil Medical Warming Catheter. The product is contained within prostate kits. This product is used in cryosurgery to prevent freezing of tissue.
Product code
GEH Unit, Cryosurgical, Accessories
Reason for recall
The warming catheter was manufactured with the inlet and outlet tubes switched.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The recalling firm telephoned their customers in April 2008 to inform them of the problem and the need to return the product. There customers were also asked to contact the firm if they had actually used any of the kits with the serial numbers listed, even if they experienced no problem. As a follow up to the telephone call, a letter was issued to the customers. For assistance, please call 770-565-6166.
Quantity in commerce
541 catheters
Distribution
Worldwide Distribution --- USA including states of PA, SC, FL, NY, TN, and WA, and country of Canada
Product codes and lots
Catalog number ASMCH0340 sold within kits with the following part numbers: FPRPR4570, FPRPR4508, FPRPR3008, FPRPR4070, FPRPR353, and FPRPR4003.