Lupine BR Anchor w/Orthocord TCP/PLGA Absorbable Anchor, size 2 (5 metric) Braided Composite Suture, violet...

DePuy Mitek, Inc. / initiated 2008-04-29 / Terminated / root cause: manufacturing process

DePuy Mitek, Inc., a Johnson & Johnson Co. began a recall of Lupine BR Anchor w/Orthocord TCP/PLGA Absorbable Anchor, size 2 (5 metric) Braided Composite Suture, violet 36" (91cm) Size 2 (5metric) eyelet loop. Made in... on . The reason FDA records is "Lupine BR Anchor w/Orthocord was assembled with an incorrect suture configuration. The suture loop was not assembled at the distal end of the anchor. It was only threaded through the suture hole in the anchor. Risk...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1810-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DePuy Mitek, Inc.
Product
Lupine BR Anchor w/Orthocord TCP/PLGA Absorbable Anchor, size 2 (5 metric) Braided Composite Suture, violet 36" (91cm) Size 2 (5metric) eyelet loop. Made in Switzerland; Sterile. The Lupine BR Anchor System is indicated for use in soft tissue bone fixation.
Product code
JDR Staple, Fixation, Bone
Reason for recall
Lupine BR Anchor w/Orthocord was assembled with an incorrect suture configuration. The suture loop was not assembled at the distal end of the anchor. It was only threaded through the suture hole in the anchor. Risk associated with incorrect suture configuration include premature breakage.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
DePuy Mitek notified customers by letter dated April 29, 2008 titled Urgent: Voluntary Medical Device Recall Letter asking consignees to isolate and do not use product and to return any of the identified product to DePuy Mitek using enclosed prepaid Federal Express shipping labels and to include a photocopy of the completed Business Reply Form. The firm is asking customers to complete the Business Reply Form with a self addressed, stamped envelope indicating level of inventory of the affected product and to return for immediate delivery of replacement product. If you have any questions contact DePuy Mitek Customer Service directly at 1-800-382-4682.
Quantity in commerce
105 units
Distribution
Class II Recall - Worldwide Distribution --- USA including states of CA, NY, MA, NV, OH, TX, TN, MD, and countries of France, UK, Canada, Spain.
Product codes and lots
Catalog Number 210711 Lot Number 3069512