Z-1824-2008
Headset Model Number RTC-551-070 for use with the Serene Sound Digital MRI Compatible High Fidelity Stereo...
Resonance Technology Inc / initiated 2007-05-14 / Terminated / root cause: design
Resonance Technology Inc began a recall of Headset Model Number RTC-551-070 for use with the Serene Sound Digital MRI Compatible High Fidelity Stereo Sound System on . The reason FDA records is "Potential safety concerns - Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to the skin". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1824-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Resonance Technology Inc
- Product
- Headset Model Number RTC-551-070 for use with the Serene Sound Digital MRI Compatible High Fidelity Stereo Sound System .
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Potential safety concerns - Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to the skin.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The strategy was notifying customers by written communication. The letter advises Resonance Technology, Inc. is voluntarily recalling the MRI Headset, Model No. RTC-2K and RTC- 551-070 because of potential safety concerns. Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to skin. The letter also advises that Resonance Technology Inc. has designed and developed the significantly improved MRI Headset, Model No. RTC-661-060-09P-001 (upgrade) and began shipping this product in January 2007. Customers who received the Headset Model; RTC-2K and RTC- 551-070, are asked to upgrade to MRI Headset Model No. RTC-661-060-09P-001. Response is requested as to the number of headsets to be replaced. In addition, consignees are advised that if they further distributed this product, to identify their customers and notify them at once of this product recall. If you have questions, contact Olivia Shayesteh at 818-882-1997 or via email at [email protected]
- Quantity in commerce
- 29 units
- Distribution
- Nationwide and Internationally: Canada, Japan, Taiwan, United Kingdom, Netherlands, Italy, Spain, Germany, Switzerland, France,
- Product codes and lots
- Model Number RTC-551-070; Serial No: a) 064-10288 b) 064-10279 c) 114-09556 d) 124-09898 e) 104-09995 f) 064-10220 g) 064-10286 h) 124-09491 i) 104-09828 j) 064-10229 k) 064-10211 l) 064-10241 m) 064-10214 n)064-09983 o) 064-10243 p) 064-10289 q) 064-10207 r) 064-10132 s) 104-09872 t) 064-10221 u) 064-10127 v) 064-10215 w) 104-09966 x) 064-10263 y) 124-10116 z) 124-09983 aa) 124-09600 bb) 114-09613 cc) 064-09479