Z-1913-2008
Boston Scientific LeVeen SuperSlim Needle Electrode, 2.0 cm/15 cm, sterile, Boston Scientific, Spencer, IN...
Boston Scientific Corporation / initiated 2008-04-30 / Terminated / root cause: design
Boston Scientific Corp began a recall of Boston Scientific LeVeen SuperSlim Needle Electrode, 2.0 cm/15 cm, sterile, Boston Scientific, Spencer, IN; REF 26-226, UPN M001262260. The product is used... on . The reason FDA records is "The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1913-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific LeVeen SuperSlim Needle Electrode, 2.0 cm/15 cm, sterile, Boston Scientific, Spencer, IN; REF 26-226, UPN M001262260. The product is used with a radiofrequency generator for the thermal coagulation necrosis of soft tissues, including partial or complete ablation of nonresectable liver lesions.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified by recall letter on 4/30/08, sent fexex priority mail, to cease use of the products and to return them to the firm. For assistance, contact Boston Scientific at 1-508-683-4186.
- Quantity in commerce
- 7,460
- Distribution
- Nationwide, Argentina, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Iraq, Ireland, Italy, Jordan, Korea, Lebanon, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, and United Arab Emirates.
- Product codes and lots
- All lots. (Lots 0007754221, 0007756536, 0007760602, 0007773373, 0007773795, 0007795354, 0007795671, 0007829946, 0007860541, 0007924564, 0007929069, 0007935435, 0007935641, 0007935927, 0007967967, 0007973448, 0007973449, 0007983095, 0007999035, 0008029591, 0008029965, 0008047351, 0008055911, 0008100038, 0008100039, 0008119923, 0008119924, 0008138446, 0008142890, 0008142891, 0008147540, 0008165769, 0008165770, 0008165771, 0008181500, 0008198428, 0008198430, 0008202143, 0008229896, 0008246451, 0008246452, 0008247161, 0008250918, 0008250919, 0008262976, 0008262977, 0008279250, 0008287295, 0008632162, 0008633030, 0008678724, 0008715160, 0008745079, 0008745080, 0008745081, 0008745084, 0008745085, 0008757525, 0008764755, 0008815843, 0008815844, 0008840208, 0008842737, 0008843687, 0008855834, 0008865948, 0008870965, 0008885317, 0008922049, 0008922050, 0008922051, 0008962320, 0008971482, 0008976593, 0008993095, 0009012204, 0009012205, 0009012206, 0009034730, 0009059144, 0009070216, 0009081512, 0009092663, 0009103629, 0009103630, 0009103631, 0009104111, 0009275204, 0009277201, 0009277202, 0009288930, 0009292225, 0009294911, 0009295272, 0009295273, 0009330954, 0009344590, 0009352769, 0009386363, 0009409068, 0009429873, 0009449403, 0009457082, 0009464931, 0009488649, 0009501969, 0009520948, 0009538053, 0009554399, 0009568323, 0009595705, 0009595853, 0009609675, 0009622354, 0009645946, 0009662038, 0009679969, 0009696743, 0009711986, 0009726408, 0009726409, 0009731711, 0009747000, 0009749812, 0009776800, 0009784043, 0009800139, 0011001910, 0011007735, 0011017381, 0011027710, 0011036114, 0011042782, 0011083691, 0011111662, 0011300194, 0011303965, 0011312752 and 0011442784.)