Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and...
Celsus Laboratories Inc began a recall of Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and in vitro diagnostic devices on . The reason FDA records is "Contamination: The firm learned that up to 2.2% oversulfated chondroitin was found in a batch of crude heparin they received from their supplier". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1917-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Celsus Laboratories Inc
- Product
- Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and in vitro diagnostic devices.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- Contamination: The firm learned that up to 2.2% oversulfated chondroitin was found in a batch of crude heparin they received from their supplier.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Celsus notified their customers by e-mail on April 24, 2008, explaining that oversulfated chondroitin was found in a batch of crude heparin they received from their supplier. Celsus recommended the device manufacturers quarantine material from the specified lots.
- Quantity in commerce
- HL-31707=11,542g, HL-31807=12,836g, HL-32708=8,374g, HL-32808=12,128g & HL-32908=5,000g
- Distribution
- Worldwide Distribution. The Lithium Heparin was sold to companies in USA (Nebraska, New Jersey, Ohio), Puerto Rico, Austria, and Denmark.
- Product codes and lots
- Lot# HL-31707, HL-31807, HL-32708, HL-32808 & HL-32908