Z-1917-2008

Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and...

Celsus Laboratories Inc / initiated 2008-03-31 / Terminated / root cause: manufacturing process

Celsus Laboratories Inc began a recall of Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and in vitro diagnostic devices on . The reason FDA records is "Contamination: The firm learned that up to 2.2% oversulfated chondroitin was found in a batch of crude heparin they received from their supplier". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1917-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Celsus Laboratories Inc
Product
Celsus Laboratories Heparin Lithium Lyophilized; Lithium Heparin for use in blood collection tube devices and in vitro diagnostic devices.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
Contamination: The firm learned that up to 2.2% oversulfated chondroitin was found in a batch of crude heparin they received from their supplier.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Celsus notified their customers by e-mail on April 24, 2008, explaining that oversulfated chondroitin was found in a batch of crude heparin they received from their supplier. Celsus recommended the device manufacturers quarantine material from the specified lots.
Quantity in commerce
HL-31707=11,542g, HL-31807=12,836g, HL-32708=8,374g, HL-32808=12,128g & HL-32908=5,000g
Distribution
Worldwide Distribution. The Lithium Heparin was sold to companies in USA (Nebraska, New Jersey, Ohio), Puerto Rico, Austria, and Denmark.
Product codes and lots
Lot# HL-31707, HL-31807, HL-32708, HL-32808 & HL-32908