Depuy LCS Knee - Orthopedic knee implant. Meniscal bearing insert, Sz: STD, mm thick 12 5, sterile; Depuy...
Depuy Orthopaedics, Inc. began a recall of Depuy LCS Knee - Orthopedic knee implant. Meniscal bearing insert, Sz: STD, mm thick 12 5, sterile; Depuy Orthopaedics, Inc., Warsaw, IN; REF 1178-21-025)... on . The reason FDA records is "Labeling Discrepancy -- Mislabeled and mis-etched as to size. Inserts were labeled and packaged as size Standard, 12.5mm Inserts but were actually a 10mm Inserts. Risks include but not limited to: a delay in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1812-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- Depuy LCS Knee - Orthopedic knee implant. Meniscal bearing insert, Sz: STD, mm thick 12 5, sterile; Depuy Orthopaedics, Inc., Warsaw, IN; REF 1178-21-025). Product is a prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing intended to to allow for motion between the plastic bearing and metal baseplate components of a knee prosthesis. It is intended to replace a knee joint in order to relieve pain and restore knee function, for indications such as osteoarthritis, inflammatory arthritis, traumatic arthritis, and revision of failed knee prostheses.
- Product code
- NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
- Reason for recall
- Labeling Discrepancy -- Mislabeled and mis-etched as to size. Inserts were labeled and packaged as size Standard, 12.5mm Inserts but were actually a 10mm Inserts. Risks include but not limited to: a delay in the procedure while the proper component is located, and tight joint with constricted movement as a result of implantation of the wrong thickness of insert.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 5/6/08 an Urgent Information - Recall Notice and an User Facility Reconciliation Report was issued to all customers/consignees informing them to find and return any product on hand and the one implanting surgeon was also notified via letter. If you have any questions contact Steve Dowell at 1-800-366-8143.
- Quantity in commerce
- 10
- Distribution
- Class II Recall - Nationwide Distribution --- including states of Pennsylvania and Virginia.
- Product codes and lots
- Lot B2LE31000.