Z-1813-2008

Abbott FreeStyle Blood Glucose Monitoring System Test Strips for use with FreeStyle Blood Glucose Monitoring...

Abbott Medical / initiated 2008-04-24 / Terminated / root cause: manufacturing process

Abbott Diabetes Care, Inc. began a recall of Abbott FreeStyle Blood Glucose Monitoring System Test Strips for use with FreeStyle Blood Glucose Monitoring Meter on . The reason FDA records is "Flaws in strip voltage continuity may result in frequency of an Error 3 ("Er3") message display and unreadable strips. This issue has the potential for creating a delay in generating blood glucose results and to delay...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1813-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott FreeStyle Blood Glucose Monitoring System Test Strips for use with FreeStyle Blood Glucose Monitoring Meter
Product code
LFR Glucose Dehydrogenase, Glucose
Reason for recall
Flaws in strip voltage continuity may result in frequency of an Error 3 ("Er3") message display and unreadable strips. This issue has the potential for creating a delay in generating blood glucose results and to delay diabetes management, potentially leading to either hypo or hyperglycemia and their associated complications.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Recall initiated on April 24, 2008. An Urgent Medical Device Correction was issued to customers. Consignees were notified, along with registered users of the FreeStyle Freedom meters, via recall letters, visits or phone calls. Distributors will be notified, along with the creation of customer lists and tracking of communications to sub-recall. If you have any questions call Tina Mazurkiewicz at 1-800-777-6565.
Quantity in commerce
2954900 strips
Distribution
Class II Recall - Worldwide Distribution --- including USA and countries of Sweden, Norway, Netherlands, Luxembourg, Switzerland, Israel and Denmark.
Product codes and lots
Part numbers/Serial numbers/Exp. Date: CAT12050-01/0618625/Jul08; M70449-03/0703325/Feb09; CAT12450-01 (70345-02)/0618129/Jun08; CAT70450-01/0700815/Jan09; CAT70447-02/0628646/Oct08; CAT12450-01 (70345-02)/0627030/Sep08; CAT70450-02/0703019/Jan09; CAT12050-02/0703118/Jan09; CAT12050-03/0715129/May09