Z-2056-2008
CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software...
Accuray Incorporated / initiated 2008-04-09 / Terminated / root cause: software
Accuray Inc began a recall of CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher on . The reason FDA records is "Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2056-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accuray Incorporated
- Product
- CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On April 16, 2008, the firm issued notifications via fedex to affected sites. In addition, installation and upgrade procedures will be updated to incorporate a step which decommissions sample beam data. If you have questions, contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or send and email to [email protected].
- Quantity in commerce
- 109 units
- Distribution
- Product was distributed to 104 medical centers throughout the US and to Vietnam, China, Korea, Taiwan, Malaysia, Turkey, France, Italy, Spain, Greece and Germany.
- Product codes and lots
- Versions 1.5.2, with system numbers (all preceded by C0) 095, 142, 058, 106, 061, 056, 128, 129, 079, 088, 116, 151, 139, 047, 070, 072, 133, 112, 149, 078, 137, 025, 098, 092, 045, 144, 124, 160, 044, 126, 123, 111, 055, 062, 074, 068, 059, 052, 103, 132, 100, 150, 051, 094, 119, 034, 146, 153, 064, 155, 037, 114, 046, 063, 097, 143, 067, 131, 042, 048, 090, 091, 122, 159, 022, 109, 099, 135, 036, 134, 127, 158, 057, 069, 030, 118, 108, 152, 136, 049, 102, 080, 101, 141, 071, 087, 140, 107, 145, 148, 120, 023, 147, 089, 038, 041, 065, 138, 060, 113, 110, 130, 043, 125, 105, 093, 117, 032, 066