Z-1853-2008

HydraGlide XL Heparin Coated Silicone Thoracic Catheters, Increased number of eyelets, Increased Length by...

Atrium Medical Corporation / initiated 2008-05-12 / Terminated / root cause: manufacturing process

Atrium Medical Corporation began a recall of HydraGlide XL Heparin Coated Silicone Thoracic Catheters, Increased number of eyelets, Increased Length by 6", Atrium Code # 14724, Size 24, # Eyelets 18, 10... on . The reason FDA records is "Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated chondroitin sulfate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1853-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Atrium Medical Corporation
Product
HydraGlide XL Heparin Coated Silicone Thoracic Catheters, Increased number of eyelets, Increased Length by 6", Atrium Code # 14724, Size 24, # Eyelets 18, 10 per case
Product code
GBZ Catheter, Cholangiography
Reason for recall
Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated chondroitin sulfate
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Atrium Medical issued a press release on 5/12/2008 and notified consignees by facsimile on 5/12/2008. Customers were instructed to contact the firm for a return goods authorization and to notify the firm upon receipt of the recall notice.
Quantity in commerce
252 cases (10 individual/case)
Distribution
Worldwide Distribution: USA, US Virgin Islands, South Africa, Austria, Taiwan, Venezuela, and Saudi Arabia.
Product codes and lots
Lot #s 10304307, 10307012, 10311125, 10311233, 10318739, 10327129, 10336323, 10336426