Z-1133-2008

GE Healthcare Definium 5000 Digital Radiographic Imaging System. Model numbers 5220493. The Definium 5000...

GE Healthcare / initiated 2007-12-17 / Terminated

GE Healthcare began a recall of GE Healthcare Definium 5000 Digital Radiographic Imaging System. Model numbers 5220493. The Definium 5000 System is designed to handle radiographic... on . The reason FDA records is "GE Definium 5000 Collimator failed compliance testing due to a blade sizing issue". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1133-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Definium 5000 Digital Radiographic Imaging System. Model numbers 5220493. The Definium 5000 System is designed to handle radiographic applications using the digital system. The system configuration includes a choice of elevating or non-elevating or no table, a floor column stand with rotating U-arm, and GE's patented Digital detector that captures radiographic images in digital form, as well as an X-ray generator/power unit.
Product code
KPR System, X-Ray, Stationary
Reason for recall
GE Definium 5000 Collimator failed compliance testing due to a blade sizing issue.
FDA root cause
Radiation Control for Health and Safety Act
Action
A consignee letter dated 12/12/07 was sent to consignees (Director or Manager of Radiological Imaging, Clinical Administrator). The letter stated that an installed collimator failed compliance testing because of a blaze sizing issue. A GE Field Engineer site visit was also conducted (FMI No. 10814 12/18/07) to implement the correction. Contact GE Healthcare at 1-262-544-3894.
Quantity in commerce
14 devices (12US & 2 OUS)
Distribution
MD, FL, AZ, TX, MS, NY, OH, CA, GERMANY, and FRANCE.
Product codes and lots
Model Number 5220493:  PC0127XR06, M1973721