ABBOTT Clinical Chemistry Phenytoin, List Number: 1E07-20, is used for the quantitation of phenytoin in human...
Abbott Laboratories Inc. began a recall of ABBOTT Clinical Chemistry Phenytoin, List Number: 1E07-20, is used for the quantitation of phenytoin in human serum or plasma on . The reason FDA records is "An Abbott investigation has determined that the phenytoin reagent does not maintain an onboard stability of 28 days as stated in the phenytoin package insert. In addition, the 14 day calibration interval is also not...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2059-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ABBOTT Clinical Chemistry Phenytoin, List Number: 1E07-20, is used for the quantitation of phenytoin in human serum or plasma.
- Product code
- DIP Enzyme Immunoassay, Diphenylhydantoin
- Reason for recall
- An Abbott investigation has determined that the phenytoin reagent does not maintain an onboard stability of 28 days as stated in the phenytoin package insert. In addition, the 14 day calibration interval is also not being met. Further investigation is required to determine the root cause of this issue. Until the cause has been determined and a corrective action defined, specific instructions ar
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- On February 15, 2008 A Product Correction letter with instrument specific attachments was provided to all Clinical Chemistry Phenytoin customers that have received the affected lots. A customer reply form was included in all US customers letter. The letter instructed customers to follow the attached instrument specific actions as described in the attachments (Attachment A: ARCHITECT cSYSTEM, Attachment B: AEROSET). Customers were also instructed to forward a copy of the Product Correction letter to any other laboratories to which they may have forwarded the affected product. If you have questions, U.S. customers should call Customer Support at 1-800-4ABBOTT and customers outside of the U.S. should contact your local customer support.
- Quantity in commerce
- 3266 Worldwide, 2009 Domestically
- Distribution
- Nationwide and Worldwide to: Australia, Canada, Chile, Germany, Hong Kong, New Zealand, Panama, Singapore, Uruguay, Singapore
- Product codes and lots
- Lot Numbers:42036HW00, 44084HW00, 45051HW00, 47055HW00, 50039HW00, 52013HW00, 55015HW00, 59088HW00