Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine...
Alphatec Spine, Inc. began a recall of Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a... on . The reason FDA records is "Firm confirmed that two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. They were incorrectly laser marked with the wrong part numbers and sizes. Two lots were...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0507-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alphatec Spine, Inc.
- Product
- Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). Intended for use with supplemental spinal fixation systems.
- Product code
- MQP Spinal Vertebral Body Replacement Device
- Reason for recall
- Firm confirmed that two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. They were incorrectly laser marked with the wrong part numbers and sizes. Two lots were mixed from the following lot numbers: Part Number 64733-120 Lot Number 611884, and Part Number 64753-126 and Lot Number 611881.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The recall was initiated on August 16, 2007 with Alphatec Spine contacting each of the four direct distributors by telephone. This was followed with a Product Recall Notice, dated August 27, 2007, that was mailed to all affected customers. The distributors were informed that the firm was conducting a voluntary recall because two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. The firm asked the customers to return the product, and were informed that replacement product will be sent back.
- Quantity in commerce
- 3 units
- Distribution
- Nationwide Distribution --- including states of AZ, FL, NY & TX.