Medtronic Extracorporeal Circuit with bio-active surface; Cardiopulmonary Bypass Oxygenator, Catalog Numbers...
Medtronic Perfusion Systems began a recall of Medtronic Extracorporeal Circuit with bio-active surface; Cardiopulmonary Bypass Oxygenator, Catalog Numbers: CB0B90R3, CB0L44R15, CB0X97R10, CB1V99R2... on . The reason FDA records is "Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1920-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Medtronic Extracorporeal Circuit with bio-active surface; Cardiopulmonary Bypass Oxygenator, Catalog Numbers: CB0B90R3, CB0L44R15, CB0X97R10, CB1V99R2, CB1Y06R14, CB1Z42R15, CB1Z69R12, CB1Z69R12, CB1Z85R16, CB2503R1, CB2K59R8, CB2K59R8, CB2S52R4, CB2U05R1, CB2W86R2, CB3E29R5, CB3L87R4, CB3U53R7, CB3X44R4, CB4741R15, CB4E05R6, CB4F32R3, CB4F32R3, CB4F32R3, CB4G10R5, CB4Q81R, CB4S04R2, CB4T12R2, CB4W67R3, CB5N65R2, HY1709R34, HY3B29R13, HY4X31R7, HY5C74R15, HY5H32R5, HY5L38R1, HY5L72R1, HY5R03R, and TL2L66R5.
- Product code
- DTZ Oxygenator, Cardiopulmonary Bypass
- Reason for recall
- Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it had provided some batches to Medtronic that were contaminated with oversulfated chondroitin sulfate (OSCS).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Medtronic issued a Press Release on 05/07/2008. An Urgent Medical Device Recall letter was sent to customers on May 07, 2008. The letter identifies affected devices, explains the situation and requests that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall Certificate acknowledging the receipt of this information.
- Quantity in commerce
- 311 US, 27 OUS
- Distribution
- Worldwide Distribution: USA, Australia, Belgium, Canada, Czech Republic, France, Italy, Japan, Norway, and Taiwan.
- Product codes and lots
- Catalog / Lot: CB0B90R3 / 9871463, CB0L44R15 / 9904942, CB0X97R10 / 9818263, CB1V99R2 / 9887322, CB1Y06R14 / 9872639, CB1Z42R15 / 9568381, CB1Z69R12 / 9805685, CB1Z69R12 / 9819156, CB1Z85R16 / 9816273, CB2503R1 / 9879856, CB2K59R8 / 9816253, CB2K59R8 / 9819175, CB2S52R4 / 9882568, CB2U05R1 / 9834608, CB2W86R2 / 9812228, CB3E29R5 / 9819173, CB3L87R4 / 9839338, CB3U53R7 / 9894969, CB3X44R4 / 9837298, CB4741R15 / 9823284, CB4E05R6 / 9892093, CB4F32R3 / 9819157, CB4F32R3 / 9819650, CB4F32R3 / 9867648, CB4G10R5 / 9885608, CB4Q81R / 9866753, CB4S04R2 / 9841997, CB4T12R2 / 9861563, CB4W67R3 / 9837302, CB5N65R2 / 9904072, HY1709R34 / 9818265, HY3B29R13 / 9895128, HY4X31R7 / 9864540, HY5C74R15 / 9838652, HY5H32R5 / 9879895, HY5L38R1 / 9859513, HY5L72R1 / 9882604, HY5R03R / 9861103, TL2L66R5 / 9859465