Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2351-2008

InterGro DBM Demineralized Bone Matrix in a lipid carrier; Provided by: Lifelink Tissue Bank, Tampa, Florida...

Michigan Orthopaedic Products, Inc. / initiated 2008-05-05 / Terminated / root cause: manufacturing process

Michigan Orthopaedic Products, Inc. began a recall of InterGro DBM Demineralized Bone Matrix in a lipid carrier; Provided by: Lifelink Tissue Bank, Tampa, Florida, EBI A Biomet Company, Parsippany, New Jersey; REF... on . The reason FDA records is "Exposure to temperature extremes -- Product was not stored under controlled conditions and may have been temperature abused. Exposure to temperature extremes could cause problems with device safety (e.g. risk of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2351-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Michigan Orthopaedic Products, Inc.
Product
InterGro DBM Demineralized Bone Matrix in a lipid carrier; Provided by: Lifelink Tissue Bank, Tampa, Florida, EBI A Biomet Company, Parsippany, New Jersey; REF DBM002. InterGro DBM product contain human tissue (allograft bpm) and are intended for transplantation. Products are to be used for filling bony voids or gaps in the extremities, and pelvis that are not intrinsic to the bony stability of the structure.
Product code
MQV Filler, Bone Void, Calcium Compound
Reason for recall
Exposure to temperature extremes -- Product was not stored under controlled conditions and may have been temperature abused. Exposure to temperature extremes could cause problems with device safety (e.g. risk of infection) or problems with device effectiveness (e.g. lower healing rates if the factors or collagen in the device were affected by the temperature extremes).
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Three units were picked up from the consignee on or about 5/05/08. The implanting physician was verbally advised of the problem with the fourth unit. If you have any questions contact Michigan Orthopaedic Products, Inc. at (616) 249-8937.
Quantity in commerce
4 packages
Distribution
Class II Recall - Nationwide Distribution --- including the state of Michigan.
Product codes and lots
Lots 259990 and 318630 (exp 07/2009).