Quantimetrix Dip and Spin Control Urine Dipstick/Microscopics Control. The intended use is to validate the...
Quantimetrix Corporation began a recall of Quantimetrix Dip and Spin Control Urine Dipstick/Microscopics Control. The intended use is to validate the performance of the Multistix, Chemstrip, Uriscan... on . The reason FDA records is "This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0514-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Quantimetrix Corp.
- Product
- Quantimetrix Dip and Spin Control Urine Dipstick/Microscopics Control. The intended use is to validate the performance of the Multistix, Chemstrip, Uriscan, QuickVue, Diascreen and other urinalysis reagent strips, and as a control for confirmatory tests such as Acetest, Clinitest, and Icotest reagent tablets, and as a control for hCG methods. In addition, the Dip and Spin Control is intended as a means of validating the processing and centrifugation of patient urine samples prior to the microscopic evaluation of urine sediment.
- Product code
- JJW Urinalysis Controls (Assayed And Unassayed)
- Reason for recall
- This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The recall communication was initiated when customer both direct and distributors who have purchased the lots were mailed a recall notification letter dated June 8, 2008 with instructions which was sent via first class mail. Customers were requested to assist in the voluntary recall of the listed product and if any of the affected lots have been consumed please review their records to ensure that the obtained values conform to the stated specifications found in the lot specific product insert.
- Quantity in commerce
- 24,937 sets for all products.
- Distribution
- Nationwide Distribution including states of AK, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NH, NM, NY, OR, PA, SC, SD, TN, TX, VA, WA and WY.
- Product codes and lots
- Lot Numbers: 47160, 47170, 47180, 47190, 47200 and 47210.