Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0515-2009

Quantimetrix SpinalScopics Spinal Fluid Cell Count Control. Spinalscopics is a plastic bottle liquid spinal...

Quantimetrix Corp. / initiated 2007-06-01 / Terminated / root cause: manufacturing process

Quantimetrix Corporation began a recall of Quantimetrix SpinalScopics Spinal Fluid Cell Count Control. Spinalscopics is a plastic bottle liquid spinal fluid cell count control. This product is intended... on . The reason FDA records is "This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0515-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Quantimetrix Corp.
Product
Quantimetrix SpinalScopics Spinal Fluid Cell Count Control. Spinalscopics is a plastic bottle liquid spinal fluid cell count control. This product is intended for monitoring cell counts in patient cerebrospinal fluid samples performed manually using a hemocytometer.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recall communication was initiated when customer both direct and distributors who have purchased the lots were mailed a recall notification letter dated June 8, 2008 with instructions which was sent via first class mail. Customers were requested to assist in the voluntary recall of the listed product and if any of the affected lots have been consumed please review their records to ensure that the obtained values conform to the stated specifications found in the lot specific product insert.
Quantity in commerce
24,937 sets for all products.
Distribution
Nationwide Distribution including states of AK, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NH, NM, NY, OR, PA, SC, SD, TN, TX, VA, WA and WY.
Product codes and lots
Lot Numbers: 46130, 46131, 46132, 46140, 46141, 46142, 46150, 46151, 46152, 46160, 46161, 46162, 46170, 46171 and 46172.