Quantimetrix Synovialscopics Synovial Fluid Control. Synovialscopics is a plastic bottle Synovial fluid cell...
Quantimetrix Corporation began a recall of Quantimetrix Synovialscopics Synovial Fluid Control. Synovialscopics is a plastic bottle Synovial fluid cell count control. It is a control for monitoring... on . The reason FDA records is "This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0516-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Quantimetrix Corp.
- Product
- Quantimetrix Synovialscopics Synovial Fluid Control. Synovialscopics is a plastic bottle Synovial fluid cell count control. It is a control for monitoring total cell counts in patient Synovial fluid samples performed manually using a hemocytometer. It is also an aid for identifying the crystals which may indicate a pathological condition present in the Synovial fluid. This control is used as a third party control to confirm proper functioning of the analysis by microscopic evaluation.
- Product code
- JPK Mixture, Hematology Quality Control
- Reason for recall
- This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The recall communication was initiated when customer both direct and distributors who have purchased the lots were mailed a recall notification letter dated June 8, 2008 with instructions which was sent via first class mail. Customers were requested to assist in the voluntary recall of the listed product and if any of the affected lots have been consumed please review their records to ensure that the obtained values conform to the stated specifications found in the lot specific product insert.
- Quantity in commerce
- 24,937 sets for all products.
- Distribution
- Nationwide Distribution including states of AK, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NH, NM, NY, OR, PA, SC, SD, TN, TX, VA, WA and WY.
- Product codes and lots
- Lot Numbers: 38050 and 38060.