Z-1864-2008

Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee...

Howmedica Osteonics Corp / initiated 2005-05-27 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee Instrumentation system. It interfaces with the Modular Handled... on . The reason FDA records is "Stryker Orthopaedics became aware that when attempting to position the Triathlon Femoral Sizer, the pre-setting may slip from l (left knee) to R (right knee) and vice verse intraoperatively". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1864-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee Instrumentation system. It interfaces with the Modular Handled and the Femoral Stylus to facilitate sizing for a femoral component, and preparing holes to accept the corresponding 4:1 Cutting Guide. The femoral sizer is designed to prepare the femoral cuts in 3 degrees of external rotation by referencing against the posterior condyles. The Femoral sizer is set by locking it into the left or right orientation based on the femur.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Stryker Orthopaedics became aware that when attempting to position the Triathlon Femoral Sizer, the pre-setting may slip from l (left knee) to R (right knee) and vice verse intraoperatively.
Root cause tag
design
FDA root cause
Device Design
Action
Letters were sent to all affected Stryker Branches/Agencies on May 27, 2005. Letters requested that each branch examine the inventory and hospitals locations to identify the instruments. Contact James Young at 201-831-5110, Brian Coughlin at 210-831-5665 or Rita Intorrella at 201-831-5825 if you have questions.
Quantity in commerce
560 units
Distribution
To Stryker branches/agencies worldwide.
Product codes and lots
Catalog Number: 6541-1-603 All lots (ecxept lot codes NMZC01, NMYV01, NMYW03.