Z-1868-2008

Encore Modular Femoral Neck, 12 degree/35 mm; Ref #410-35-112; distributed by Encore Medical, Austin, TX...

Encore Medical, LP / initiated 2008-05-08 / Terminated / root cause: manufacturing process

Encore Medical, Lp began a recall of Encore Modular Femoral Neck, 12 degree/35 mm; Ref #410-35-112; distributed by Encore Medical, Austin, TX 78758. Designed to ensure the proper anteversion and... on . The reason FDA records is "Product not manufactured to specifications; modular femoral necks found to contain 24 scallops, instead of the standard 12". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1868-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Encore Modular Femoral Neck, 12 degree/35 mm; Ref #410-35-112; distributed by Encore Medical, Austin, TX 78758. Designed to ensure the proper anteversion and femoral head orientation in the acetabulum.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Product not manufactured to specifications; modular femoral necks found to contain 24 scallops, instead of the standard 12.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Firm notified sales agents of recall on 05/08/08 via an Urgent: Medical Device Recall letter. Sales agents asked to immediately examine their inventory and quarantine product subject to recall. Agents are asked to notify their customers if they had further distributed the product. All product must be returned to Encore along with the response form.
Quantity in commerce
20 units
Distribution
Worldwide Distribution --- USA including states of IN, KS, GA, WA, SD, and countries of Greece and Australia.
Product codes and lots
Lots #53876569, Exp 7/23/2013; 53869022, and lot 53873037