Z-2151-2008

Equal Diagnostics Lipase Color Reagent Kit Part Number: 905-E For In Vitro Diagnostic Use Containing the...

Genzyme Corporation / initiated 2008-04-14 / Terminated / root cause: manufacturing process

Genzyme Corporation began a recall of Equal Diagnostics Lipase Color Reagent Kit Part Number: 905-E For In Vitro Diagnostic Use Containing the Lipase Color Reagent, 30 ml, Part Number 2005-01, lot... on . The reason FDA records is "Calibration and Quality Control outside of established range resulting from an increase in initial OD readings with the reagent contained within the kits". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2151-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Genzyme Corporation
Product
Equal Diagnostics Lipase Color Reagent Kit Part Number: 905-E For In Vitro Diagnostic Use Containing the Lipase Color Reagent, 30 ml, Part Number 2005-01, lot Number 491686 .
Product code
CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
Reason for recall
Calibration and Quality Control outside of established range resulting from an increase in initial OD readings with the reagent contained within the kits.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Genzyme notified accounts by Fed'X on 4/14/08 via an Imprtant Field Correction letter, requesting users to discard kits and follow the "User Update" instructions previously provided to minimize calibration error. Customers are to complere and sign the Confirmation of Notification form and fax it back to the firm regardless if they have any affected product or not. Also, requires that the customer confirms that the REagent OD limits for First and Last H are set at 0.15 for Lipase on the Olympus AU400/400e, AU600, AU640/640e and AU2700/AU5400 Systems.
Quantity in commerce
3,443
Distribution
Nationwide Distribution
Product codes and lots
Reagent lot Number 491686, Lot #7EPP31 exp 01/15/09