Z-2093-2008

Boston Scientific LeVeen SuperSlim Needle Electrode, 3.0 cm/15 cm, sterile. The device is intended to be used...

Boston Scientific Corporation / initiated 2007-10-01 / Terminated / root cause: design

Boston Scientific Corp began a recall of Boston Scientific LeVeen SuperSlim Needle Electrode, 3.0 cm/15 cm, sterile. The device is intended to be used in conjunction with a Radio Therapeutics... on . The reason FDA records is "The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2093-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific LeVeen SuperSlim Needle Electrode, 3.0 cm/15 cm, sterile. The device is intended to be used in conjunction with a Radio Therapeutics Corporation radiofrequency (RF) generator for thermal coagulation necrosis of soft tissue, including partial or complete ablation of nonresectable liver lesions.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.
Root cause tag
design
FDA root cause
Device Design
Action
Customer notifications were sent to customers in Japan on October 1, 2007 instructing them to check their inventory and to return affected product. Contact Rob Miragliuolo at 508-683-4186 if you have questions.
Quantity in commerce
15,604 of all products
Distribution
Japan.
Product codes and lots
All lots.