ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected...
Abbott Laboratories, Inc began a recall of ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected software: v2.20, v2.20DB, v2.60, v3.10, 3.11 and 3.12... on . The reason FDA records is "Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2231-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected software: v2.20, v2.20DB, v2.60, v3.10, 3.11 and 3.12; distributed by Abbott Laboratories, Irving, TX. The Abbott ARCHITECT cSystem is designed to perform automated, random access, clinical chemistry analyzer which utilizes spectrophotometry (monochromatic and bichromatic modes of measurement) for photometric based determinations.
- Product code
- CGZ Electrode, Ion-Specific, Chloride
- Reason for recall
- Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Firm notified all ARCHITECT c8000 customers with a correction letter and response form distributed on 5/23/2008. The letter provided actions the users could take until an updated software version was available. For assistance, contact Abbott at 1-877-4ABBOTT.
- Quantity in commerce
- 2,378 units worldwide (431 in the USA; plus 1,947 International)
- Distribution
- Worldwide distribution, including: USA (including Puerto Rico), Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, France, Germany, Grand Cayman, Greece, Guatemala, Honduras, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Uruguay, and Venezuela.
- Product codes and lots
- Version 2.20, List #05F48-13, Part #7-201738-01; Version 2.20DB, List #05F48-14, Part #7-201738-02; Version 2.60, List #05F48-17, Part #7-203715-01; Version 3.10, List #05F48-18, Part #7-203715-02; Version 3.11, List #05F48-20, Part 7-203715-03; and Version 3.12, List #05F48-21