ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected...

Abbott Medical / initiated 2008-05-23 / Terminated / root cause: software

Abbott Laboratories, Inc began a recall of ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected software: v2.20, v2.20DB, v2.60, v3.10, 3.11 and 3.12... on . The reason FDA records is "Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2231-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected software: v2.20, v2.20DB, v2.60, v3.10, 3.11 and 3.12; distributed by Abbott Laboratories, Irving, TX. The Abbott ARCHITECT cSystem is designed to perform automated, random access, clinical chemistry analyzer which utilizes spectrophotometry (monochromatic and bichromatic modes of measurement) for photometric based determinations.
Product code
CGZ Electrode, Ion-Specific, Chloride
Reason for recall
Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample.
Root cause tag
software
FDA root cause
Software design
Action
Firm notified all ARCHITECT c8000 customers with a correction letter and response form distributed on 5/23/2008. The letter provided actions the users could take until an updated software version was available. For assistance, contact Abbott at 1-877-4ABBOTT.
Quantity in commerce
2,378 units worldwide (431 in the USA; plus 1,947 International)
Distribution
Worldwide distribution, including: USA (including Puerto Rico), Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, France, Germany, Grand Cayman, Greece, Guatemala, Honduras, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Uruguay, and Venezuela.
Product codes and lots
Version 2.20, List #05F48-13, Part #7-201738-01; Version 2.20DB, List #05F48-14, Part #7-201738-02; Version 2.60, List #05F48-17, Part #7-203715-01; Version 3.10, List #05F48-18, Part #7-203715-02; Version 3.11, List #05F48-20, Part 7-203715-03; and Version 3.12, List #05F48-21