Artisan Control Catheter, model number 04454, manufactured by Hansen Medical, Inc., Mountain View, CA...
Hansen Medical Inc began a recall of Artisan Control Catheter, model number 04454, manufactured by Hansen Medical, Inc., Mountain View, CA. Steerable guide catheter for human use intended to... on . The reason FDA records is "The catheter's flexible bellows portion may develop a leak. This has the potential to cause loss of hemostasis, flush fluid leakage, and/or introduction of air into the catheter with a risk of subsequent embolism". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2106-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hansen Medical, Inc.
- Product
- Artisan Control Catheter, model number 04454, manufactured by Hansen Medical, Inc., Mountain View, CA. Steerable guide catheter for human use intended to facilitate manipulation, positioning and control for collecting electrophysiological data within the heart atria.
- Product code
- DRA Catheter, Steerable
- Reason for recall
- The catheter's flexible bellows portion may develop a leak. This has the potential to cause loss of hemostasis, flush fluid leakage, and/or introduction of air into the catheter with a risk of subsequent embolism.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On May 21, 2008 an Urgent: Voluntary Product Recall Notification was issued. All customers who received affected product were verbally contacted directly by the firm, and were sent a follow up letter, asking them to segregate the product. Replacements will be shipped to them. If you have questions please contact Doug Worth at 1-650-404-5940.
- Quantity in commerce
- 369 units
- Distribution
- Class II Recall - Worldwide Distribution ---- including USA and countries of UK, Italy, Germany, and Czech Republic.
- Product codes and lots
- Lot numbers: 2182, 2185, 2186, 2199, 2200-2203, 2205, 2206, 2008, 2213-2218, 2220, 2223-2231, 2233, 2234, 2236-2245, 2247, 2249, 2251, 2261-2263, 2265, 2267-2271, 2273, 2274, 2276-2279, 2284, 2288-2297, 2301-2303, 2305, 2309, 2315, 2318, 2320-2322, 2324, 2327, 2328, 2330, 2332, 2334, 2342, 2344-2349, 2351, 2353, 2355-2357, 2359-2363, 2366-2367, 2370, 2370, 2374-2375, 2377-2378, 2381-2382, 2384, 2387-2389, 2391, 2392, 2394,2395, 2397, 2399, 2400, 2404, 2405, 2412, 2417, 2420, 2421, 2423-2426, 2429, 2430, 2432-2434, 2436, 2442, 2444, 2446-2448, 2450, 2453, 2457-2461, 2463-2467, 2469-2472, 2475-2483, 2485-2490, 2494, 2497, 2500, 2502-2504, 2508, 2510, 2512-2518, 2522, 2526, 2529, 2531-2532, 2534, 2536, 2538, 2539, 2547, 2549, 2551, 2553, 2556, 2558-2560, 2562-2565, 2568, 2569, 2571, 2574-2577, 2579, 2581-2582, 2586, 2588-2589, 2591-2596, 2601, 2604, 2609, 2610, 2612, 2617, 2619, 2626, 2627, 2636, 2637, 2639-2642, 2644, 2675, 2682, 2684, 2685, 2688, 2689, 2698-2699, 2701-2707, 2709, 2716, 2720, 2722-2723, 2726-2727, 2729, 2732-2735, 2738, 2741, 2743, 2745, 2746, 2752, 2754-2757, 2761, 2767, 2773-2774, 2776, 2779, 2782, 2783, 2785-2789, 2791, 2797-2798, 2801, 2803-2804, 2806, 2809, 2812, 2814-2815, 2817-2818, 2821, 2822, 2824, 2827, 2829-2831, 2833-2837, 2839, 2843, 2848, 2850, 2854, 2857, 2862, 2865, 2872, 2879, 2882, 2886, 2888, 2889, 2894, 2896, 2898, 2904, 2908-2909, 2913, 2915, 2920, 2922, 2924, 2925, 2930, 2932, and 2940.