Z-2134-2008

Vacuette, Plastic Cannula HOLDEX, Ref 450216, Single packed, sterile, single use disposable tube holder with...

Greiner Bio-One North America, Inc. / initiated 2008-05-16 / Terminated / root cause: design

Greiner Bio-One North America, Inc. began a recall of Vacuette, Plastic Cannula HOLDEX, Ref 450216, Single packed, sterile, single use disposable tube holder with off center plastic cannula at the top and a... on . The reason FDA records is "At removal of the tube from the holder the needle may dislodge and become stuck in the tube stopper with the blunt end of the needle facing out". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2134-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Greiner Bio-One North America, Inc.
Product
Vacuette, Plastic Cannula HOLDEX, Ref 450216, Single packed, sterile, single use disposable tube holder with off center plastic cannula at the top and a stainless steel needle on the underside of the holder. It is intended for use in routine specimen withdrawal from collection bags or bottles through needless cannula ports.
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
At removal of the tube from the holder the needle may dislodge and become stuck in the tube stopper with the blunt end of the needle facing out.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified by letter on/about 05/16/2008. They were instructed to stop distributing the affected lots immediately and destroy any of the product they may have. A Product Disposition form was attached to be completed and returned via fax to Greiner bio-one. Consignees instructed to notify their customers and to provide Greiner bio-one with a list of their customers that have received the item. If you have any questions contact Manfred Abel at 704-261-7823.
Quantity in commerce
1,606,005 units
Distribution
Class II Recall - Nationwide Distribution.
Product codes and lots
Item #450216, Lot #A10074, A100745, A100746, A100747, A110749, and A110750.