Z-2136-2008
Drystar AXYS, Hardcopy Printer A free standing device used to print diagnostic conventional and mammography...
AGFA Healthcare Corp. / initiated 2008-05-16 / Terminated / root cause: manufacturing process
AGFA Corp. began a recall of Drystar AXYS, Hardcopy Printer A free standing device used to print diagnostic conventional and mammography images on transparent film for viewing on a... on . The reason FDA records is "The "Film Calibration" setting on the printers was set to the default "OFF" position instead of "ON"". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2136-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AGFA Healthcare Corp.
- Product
- Drystar AXYS, Hardcopy Printer A free standing device used to print diagnostic conventional and mammography images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable.
- Product code
- LMC Camera, Multi Format, Radiological
- Reason for recall
- The "Film Calibration" setting on the printers was set to the default "OFF" position instead of "ON".
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Consignees were notified by an Urgent Safety Notice letter on/about 05/16/2008. The letter provides the procedure the customer can follow to change the film calibration setting from "OFF" to "ON". The letter also provides information on how to get assistance from Agfa, if the customers prefer a service engineer perform the change. Consignees were instructed to send a copy of the letter or similar letter to their affected accounts. Agfa also offers to dispatch a service engineer to perform the parameter change.
- Quantity in commerce
- 33 units
- Distribution
- Worldwide Distribution --- USA including states of DE, FL, IL, KS, MA, MD, ME, MO, MT, NJ, NY, OH, OK, TX, UT, VA, WA, WV, and Puerto Rico and country of Canada.
- Product codes and lots
- ID number EYZ4E, Serial numbers: 1024 to 1189; but excluding 1053, 1057, 1089, 1096, 1097, 1098, 1101, 1103, 1161, 1166, 1173, 1176, 1177, 1178, 1180, 1181, 1185, 1186, 1187 and 1188.